Storage and transport
humidity
This symbol indicates the
acceptable humidity range for
storage and transport
environments.
Storage and transport
pressure
This symbol indicates the
acceptable atmospheric
pressure range for storage and
transport environments.
For accessories
This symbol indicates the
batch code.
This symbol indicates the do
not reuse.
For customers in the U.S.A.
This equipment has been tested and found
to comply with the limits for a Class A
digital device, pursuant to part 15 of the
FCC Rules. These limits are designed to
provide reasonable protection against
harmful interference when the equipment is
operated in a commercial environment.
This equipment generates, uses and can
radiate radio frequency energy and, if not
installed and used in accordance with the
instruction manual, may cause harmful
interference to radio communications.
Operation of this equipment in a residential
area is likely to cause harmful interference
in which case the user will be required to
correct the interference at his own expense.
You are cautioned that any changes or
modifications not expressly approved in
this manual could void your authority to
operate this equipment.
All interface cables used to connect
peripherals must be shielded in order to
comply with the limits for a digital device
pursuant to Subpart B of part 15 of FCC
Rules.
For customers in the U.S.A.
Caution
Federal law (United States of America)
restricts this device to sale by or on the
order of a licensed healthcare practitioner.

For customers in Canada

This unit has been certified according to
Standard CAN/CSA-C22.2 No. 60601-1.
Important safeguards and
notices for use in the
medical environments
1. All devices connected to the unit must
be certified or compliant according to
IEC 60601-1, IEC 60950-1, and
IEC 60065 standards and other IEC/
ISO standards applicable to the devices.
2. Furthermore, the system as a whole
must comply with IEC 60601-1
standards. All peripheral devices
connected to the signal input/output
sections of the unit constitute the
medical-use system, and therefore, the
user is responsible for ensuring that the
system as a whole complies with
IEC 60601-1 standards. If in doubt,
consult qualified Sony service
personnel.
3. Connecting the unit to other devices
may increase the leakage current.
4. For all peripheral devices connected to
the unit that operate on commercial
power supplies and do not comply with
IEC 60601-1 standards, incorporate an
isolation transformer that complies with
IEC 60601-1 standards and connect to
the commercial power supply via the
transformer.
GB
3