Safety instructions:
Zone 0

Temperature tables

Connection data

Overvoltage protection

• If an overvoltage protection according to IEC/EN 60079-14 against atmospheric over voltages is
required: no other circuits may leave the housing during normal operation without additional
measures.
• For installations which require overvoltage protection to comply with national regulations or
standards (e.g. IEC/EN 60079-14), install the device using overvoltage protection (e.g. HAW56x
from Endress+Hauser).
• Observe the safety instructions of the overvoltage protection.
• In the event of potentially explosive vapor/air mixtures, only operate the device under atmospheric
conditions.
– Temperature: –20 to +60 °C
– Pressure: 80 to 110 kPA (0.8 to 1.1 bar)
– Air with normal oxygen content, usually 21 % (V/V)
If no potentially explosive mixtures are present, or if additional protective measures have been taken
according to EN 1127-1, the device may also be operated under non-atmospheric conditions in
accordance with the manufacturer' s specifications.
• Associated devices with galvanic isolation between the intrinsically safe and non-intrinsically safe
circuits are preferred.
• If there is a risk of dangerous potential differences within Zone 0 (e.g. through the occurrence of
atmospheric electricity), implement suitable measures for intrinsically safe circuits in Zone 0 (e.g. in
accordance with the requirements of IEC/EN 60079-14).
→  35
Basic specification, Position 1, 2 (Approval) = B4, I4
Ex ia
• Power supply and signal circuit with protection type: intrinsic safety Ex ia IIC or IIB
Basic specification, Position 3 (Power Supply; Output) = E, G
Terminal 1 (+), 2 (–)
FISCO:
Entity:
U
= 17.5 V
U
= 30 V
i
i
300 mA
I
= 550 mA
I
i
i
P
= 5.5 W
P
= 1.2 W
i
i
effective inner inductance L
= 10 μH
i
effective inner capacitance C
= 5 nF
i
Service interface (CDI)
Taking the following values into consideration, the device can be connected to the certified
Endress+Hauser FXA291 service tool or a similar interface:
Service interface
U
= 7.3 V
i
effective inner inductance L
= negligible
i
effective inner capacitance C
= negligible
i
U
= 7.3 V
o
I
= 100 mA
o
P
= 160 mW
o
L
(mH) =
5.00
2.00
1.00
o
C
(μF) =
0.73
1.20
1.60
o
Terminal 3 (+), 4 (–)
Switch output (PFS):
U
= 30 V
i
I
= 300 mA
i
P
= 1 W
i
effective inner inductance L
effective inner capacitance C
effective inner capacitance to ground C
0.50
0.20
0.10
0.05
2.00
2.60
3.20
4.00
XA00689F-A
TRC [26]
= 0 mH
i
= 3 nF
i
= 5.28 nF
i
0.02
0.01
0.005
0.002
5.50
7.30
10.00
12.70
0.001
12.70
21