Herunterladen Inhalt Inhalt Diese Seite drucken

A&D BA Serie Schnellstartanleitung Seite 16

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 35
EN
Display
Code
EC,E00
EC,E01
EC,E02
EC,E03
EC,E04
EC,E06
EC,E07
7.1. Other display
This is the automatic sensitivity adjustment notice (the
If the balance is not used for several minutes with this mark blinking, the balance
automatically performs sensitivity adjustment using the internal mass. (The blinking
period depends on the operating environment.)
Advice
Although it is possible to continue using the balance even while this mark is blinking,
use after sensitivity adjustment is completed is advisable in order to maintain the
weighing accuracy.
Communications error
A protocol error occurred in communications.
Check the format, baud rate, and parity.
Undefined command error
An undefined command was found.
Check the transmitted command.
Not ready
The received command cannot be executed:
(e.g.) Q command was received when not in the weighing mode.
(e.g.) Q command was received while rezeroing.
Adjust the delay time to transmit a command.
Timeout error
With the timeout parameter set to
approximately 1 second or more while receiving command characters.
Check the communication.
Character length error
The number of characters in the received command has exceeded the
limit. Check the command to transmit.
Format error
The description of the received command is incorrect:
(e.g.) The number of digits of numerical values is incorrect.
(e.g.) There are alphabet characters among the numerical values.
Check the transmitted command.
Parameter setting error
The value of the received command has exceeded the allowed value.
Check the setting range of the numerical value of the command.
Description
16
, there was a wait time of
mark blinking).

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis