SAA
VET control kit
For veterinary use only!
Intended use:
For quality control procedures of the quantitative determination with the
SAA VET test kit on a solo or CUBE-VET analyser
English
Order information
Order number: VT0170
i
Order number: VT0171
Order number: VT1700
Test kit preparation: Allow single test at least 10 minutes to warm up to room temperature (20 - 25 °C) by placing the test into the
Y
test kit rack. Put test kit package back into refrigerator.
Summary
This SAA VET control kit is intended for use as a test sample of known concentration for the SAA VET assay . The user
can compare observations with expected ranges as a means of assuring consistent performance of the solo SAA VET
test kit. The SAA VET control is prepared from a human serum base.
Required Material
(not supplied with the kit)
- Pipettes
- Distilled water
Preparation and Handling
Control material is lyophilized and must be reconstituted
before use!
Open the vial very carefully, avoiding any loss of
1.
the lyophilized material.
Add exactly 2 ml of distilled/deionized water.
2.
Warning: Inaccurate reconstitution of the control
and error in assay technique can cause erroneous
results.
Close the vial carefully and dissolve the compo-
3.
nents completely by occasional gently swirling.
Avoid the formation of foam.
Allow the contents to reconstitute over 15 minutes
4.
before.
Analyse the control in the same way as a patient
5.
sample.
Stability and Storage
Unopened vials are stable at 2 - 8 °C until the expiration
date stated on the label.
Stability after reconstitution:
at 2 - 10 °C:
Close immediately after use. Store control tightly capped
when not in use. Bacterial contamination produces an in-
crease in turbidity and/or a characteristic odor. Discard
vial if evidence of microbial contamination is observed.
Waste Management
Please refer to local legal requirements
Indication
SAA VET test kit
SAA VET test kit
SAA VET control kit
2 weeks
Eurolyser Diagnostica GmbH
Bindergasse 3
5020 Salzburg/Austria
Warnings and Precautions
This control kit is for veterinary diagnostic use only!
DO NOT INGEST! Excercise the normal precautions
required for handling all laboratory reagents. This control
utilizes human origin serum. The serum is negative HBs
antigen, HIV (HIV -1, HIV- 2) and HCV antibodies test. It is
impossible to completely rule out the possibility that the-
se viruses or other pathogens may be present. Therefore
when using these reagents, handle them with same de-
gree of caution as patient serum.
This control must be handled in accordance with recom-
mendations from Centers for Disease Control/National
Institutes of Health manual, "Biosafety in microbiological
and biomedical laboratories" (1999).
Y
Set the species to „Control" before running the
QC material
Y
Compare the results with the values indicated in
the lot-specific value table
If the result is not within the acceptable range,
Y
repeat measurement. If result is again not within
the acceptable range, please contact your distri-
butor
Tel. +43 (0)662 / 43 21 00
Fax +43 (0)662 / 43 21 00-50
www.eurolyser.com
Kit size
6 tests
16 tests
1 x 2 ml (decision level)