Description Of The Equipment - CBM 2429/V Betriebsanleitung

Heissluft inkubator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

C.B.M. S.r.l. Medical Equipment - Via Castello 10, 26038 Torre dé Picenardi (CR) – Italy
Remarks:
These instructions have to be carefully kept near the equipment, in order to be consulted by all users.
In case of lack of voltage or lighting of the alarm pilot lamp the equipment stops the incubation cycle.
We recommend not to lean any object on the equipment.
Temperature oscillations of few ° C, during the cycle, have to be considered normal.
Always place one of the perforated grids on the lower plane (1).

Description of the equipment :

Referring to the figure of page 2 the incubator components are described here below:
• (A) Master switch
• (B) "Ready machine" yellow pilot lamp
This lamp is on when the incubator is ready and it is waiting the start of a cycle
(C) Alarm red lamp
This pilot lamp turns on in order to signal some possible anomalies and it is matched to a BIP sound.
To set aside the alarm push the START button (see Alarms Table).
(D) START push button
The function of this push button is to start a cycle, if the pilot lamp (B) is turned on.
The START push button works also as RESET key, if it is pressed during an incubation cycle the cycle is
cancelled. Moreover, if an alarm has intervened, by pressing the START push button, the alarm is
cancelled.
(E) Switch push button
The function of this push button is to alternate, on the display (H), the visualization of temperature and
incubation time.
(F) Temperature pilot lamp
When this pilot lamp is turned on it shows that the value read and/or set on the display (H) is referred to
the temperature.
(G) time pilot lamp
When this pilot lamp is turned on it shows that the value read and/or set on the display (H) is referred to
the incubation duration, heating excluded.
If it flashes it means that INFINITE time has been set.
(H) display
Cycle parameters are visualized on the display : temperature inside the chamber, incubation time, set
temperature.
In order to reset the cycle data, this must be stopped.
The temperature can be set between 30° C and 70° C,
Time can be set between 1 minute and 600 minutes or infinite time.
To set the INFINITE time the push button
wait few seconds.
When the writing "INF" appear on the display wait 3 seconds for automatic storage.
When INFINITE time is selected the Lamp Time (G) flashes.
To remove INFINITE time press the button
ATTENTION: AT INFINITE TIME THE EQUIPMENT CAN BE SWITCHED OFF ONLY MANUALLY.
Rev.3 – 28/07/2010
DRY INCUBATOR- USER MANUAL
ó
½
+
must be kept pressed up to 600 and then it is necessary to
2429-2430-2431-V COMP- r3x4.DOC
.
pag. 6 di 17