  • Seite 2 Schaltstift Switch pin Schaltstiftverlängerung Switch pin extension Adapterkappe Adapter Generatorzelle Generator cell Stromversorgung intern Internal power supply LED grün LED green LED rot LED red Feuchthaltekappe Humidity cap Nur bei COCl COCl only: Anschlußbuchse External battery connector 10 Stromversorgung External power supply Nur bei COCl und SO COCl...
  • Seite 3: Inhaltsverzeichnis

    Inhaltsverzeichnis / Table of Contents Seite / Page Deutsch 1. Bedienungsanleitung 1.1 Funktionsbeschreibung 1.2 Testablauf 2. Wartung 2.1 Batteriewechsel 2.2 Austausch der Generatorzelle 2.2.1 H S und HCN Generator 2.2.2 CO, Cl und NO Generator 2.2.3 COCl und SO Generator English 3.
  • Seite 4: Funktionsbeschreibung

    Bedienungsanleitung Funktionsbeschreibung Der Compur Gasgenerator ermöglicht die Überprüfung von Gasdetektoren. Das Gerät ist mit einer gasspezifischen Generatorzelle bestückt. Je nach Gasart wird entweder elektrochemisch oder thermisch eine über dem MAK -Wert liegende Gaskonzentration erzeugt. Mit dem erzeugten Gas wird der Gasdetektor auf sicheres Ansprechen getestet. Der Gasgenerator ist nicht geeignet, ein Prüfgas definierter Konzentration zu erzeugen.
  • Seite 5: Wartung

    Wartung Sollte eine Störung an Ihrem Gasgenerator auftreten, so wenden Sie sich bitte an den Compur Monitors Kundendienst in München oder an Ihren lokalen Repräsentanten. Kleinere Wartungsarbeiten wie den Batteriewechsel oder Austausch der Generatorzelle können Sie selbst vornehmen. Beachten Sie bitte dazu folgende Hinweise: Führen Sie die Arbeiten möglichst auf einer leitfähigen, geerdeten Arbeitsmatte aus.
  • Seite 6: Austausch Der Generatorzelle

    Austausch der Generatorzelle Die Lebensdauer der Generatorzelle beträgt ca. ein Jahr bzw. 2.500 Tests. Danach sollte sie ausgetauscht werden. Die Generatorzelle sollte auch in jedem Fall getauscht werden, wenn ein nachweislich intakter Gasdetektor beim Funktionstest nicht mehr anspricht. 2.2.1 S und HCN Generator (siehe Abbildung 1) 1 Lösen Sie die vier Schrauben an der Rückseite des Gerätes und heben Sie das Gerätevorderteil ab.
  • Seite 7: Generator (Siehe Abbildung 2)

    2.2.2 CO, Cl und NO Generator (siehe Abbildung 2) Diese Generatorzellen enthalten einen Lüfter, der fest mit der Generatorzelle verbunden ist. Beim Austausch verfahren Sie wie folgt: 1 Lösen Sie die vier Schrauben an der Rückseite des Gerätes und heben Sie das Gerätevorderteil ab.
  • Seite 8: Generator (Siehe Abbildung 3)

    2.2.3 COCl und SO Generator (siehe Abbildung 3) 1 Stecken Sie den externen Batteriebehälter aus (nur bei COCl 2 Lösen Sie die vier Schrauben an der Rückseite des Gerätes und heben Sie das Gerätevorderteil ab. 3 Entnehmen Sie die Batterie. 4 Schieben Sie die beiden Isolierschläuche zurück und löten Sie die Anschlussdrähte der Generatorzelle von den Litzen.
  • Seite 9: Operating Instructions

    • When the detector is pressed down, the generation of gas starts automatically. If you want to test the Compur Monitors gas detectors Monitox plus or Dositox, please use the switch pin extension included with the generator. Compur gas detectors Monitox S/SD and Minitox are tested without extension.
  • Seite 10: Maintenance

    Maintenance In service is needed, please contact Compur Monitors Service in Munich or your local representative. Minor maintenance work such as battery replacement, or generator cell exchange, can be performed by observing the following instructions. Please observe the following notes: Use a grounded conductive pad.
  • Seite 11: Generator Cell Replacement

    Generator Cell Replacement The generator cell has an approximate life time of 1 year or 2,500 tests. After this time it needs replacement. It also needs replacement, if a gas detector known to be good does not respond to a gas test. 4.2.1 S and HCN Generator (see Picture 1) 1 Loosen the 4 screws on the backside of the instrument and separate the two halves.
  • Seite 12: Co, Cl And No Generator

    4.2.2 CO, Cl and NO Generator (see Picture 2) The CO, Cl and NO generator cells contain an internal fan. When replacing, please proceed as follows: 1 Loosen the 4 screws on the backside of the instrument and seperate the two halves. (Use caution as the LED’s are attached to the top case) 2 Remove the battery.
  • Seite 13: Cocl And So Generator

    4.2.3 COCl and SO Generator (see Picture 3) 1 Unplug the external battery (COCl only). 2 Loosen the 4 screws on the backside of the instrument and separate the two halves. (Use caution as the LED’s are attached to the top case) 3 Remove the battery.
  • Seite 14: Ersatzteile /Spare Parts

    Ersatzteile / Spare Parts Generatorzellen / Generator cells Bestell-Nr.: Article-No. 516722 COCl 516847 HCN 100 ppm 516730 HCN 20 ppm 516850 516912 516938 516755 516763 Technische Daten / Technical Data Abmessungen / Dimensions 133 x 65 x 40 mm Gewicht mit Batterie / Weight incl. battery 250 g Temperaturbereich / Temperature range +10°C - +45°C...
  • Seite 15: Abbildungen / Drawings

    Abbildungen / Drawings S, HCN Generator Abbildung / Drawing 1 Downloaded from www.Manualslib.com manuals search engine...
  • Seite 16 CO, Cl , NO Generator Abbildung / Drawing 2 Downloaded from www.Manualslib.com manuals search engine...
  • Seite 17: Cocl 2 And So 2 Generator

    COCl , SO Generator Abbildung / Drawing 3 Downloaded from www.Manualslib.com manuals search engine...