Operating the Valve Drive S 6/-12/-16
The valve drive S 16, exclusively used to control the 17-port/1-channel
switching valve, can only be operated with the DIP 1 switch in the OFF
position!
Using DIP 2 the control mode is set to either FORWARD/ BACKWARD/HOME or
to the absolute position input. This will simultaneously activate the inputs BIN 0
through BIN 2.
The connector OUT/BIN3 can be used either as output or as input. The selection is
performed by setting DIP 3 and DIP 4.
The DIP-switches 3 and 4 always should be set in opposite positions.
Not all of the possible combinations of DIP positions are permitted when operating
the valve drive. If an improper DIP setting is used, an error message will appear
either when switching the drive on or when the valve is switched.
The DIP setting always has to be done with the drive switched off. Otherwise
the DIP positions will not become active.
Table 2 DIP-switch positions in valve drive S 6
DIP 1
DIP 2
DIP 3
DIP 4
DIP 3
DIP 4
Table 3 DIP-switch positions in valve drive S 12
DIP 1
DIP 2
DIP 3
DIP 4
DIP 3
DIP 4
ON
360° valve (7-p/1-ch)
OFF
Two position valves (6-p/2-ch and 6-p/3-ch)
ON
binary control BIN 0-2 (360° valves only)
OFF
RS-232 control/ manually/
HOME/ INJ./ LOAD
ON
OUT/ BIN 3 output
OFF
OFF
not allowed
ON
ON
360° valve (13-p/1-ch)
OFF
not allowed
ON
binary control BIN 0-2
OFF
RS-232 control/ manually/
HOME/ INJ./ LOAD
ON
OUT/ BIN 3 output
OFF
OFF
OUT/ BIN 3 input
ON