DEVICE DESCRIPTION
®
®
The Mityvac
Mystic
II M-Style
®
Mushroom
Cup is designed to
assist a practitioner in the delivery of
an infant during childbirth. The
Mityvac Mystic II is an integrated pump and cup system, packaged,
and sterilized for immediate use.
WARNINGS
DO NOT exceed recommended vacuum levels.
Never apply the vacuum cup to any portion of the
infant's face.
Vacuum-assisted delivery should only be performed or
supervised by a trained and experienced operator.
 "There should be a willingness to abandon attempts at
vacuum extraction if satisfactory progress is not made."
(ACOG Technical Bulletin #196, Aug. 1994)
Abandon vacuum-assisted delivery if:
a. Vacuum cup becomes disengaged (pops off) three times
b. Vertex has not advanced substantially with each
traction attempt
c. There is evidence of fetal scalp trauma
d. Cumulative traction time exceeds 10 minutes or total
vacuum procedure time exceeds 15 to 30 minutes
If the extractor cup becomes disengaged, check for trauma
to the fetal scalp before reapplying.
CAUTION: U.S. Federal law restricts this device to sale by or on the
order of a physician.
INDICATIONS FOR VACUUM-ASSISTED VAGINAL DELIVERY
No indication is absolute.
Fetal:
Non-reassuring fetal status
Failure to deliver spontaneously following an appropriately
managed second stage of labor
Maternal:
Need to avoid voluntary expulsive efforts
Inadequate expulsive effort
CONTRAINDICATIONS
Do not initiate vacuum if any of the following conditions exist:
Non-vertex positions (breech or transverse lie/position) or face
or brow presentation
Suspected cephalopelvic disproportion
Previous scalp sampling
Suspected macrosomia, or risk of shoulder dystocia
Failed vacuum or forceps attempt
Less than 34 weeks gestation
Unengaged vertex
Incompletely dilated cervix
Need for active device rotation
Suspected fetal bleeding abnormalities
ADVERSE EVENTS
Fetal Injuries: Head trauma, bruises, contusions, lacerations,
scalp edema, skull fracture, cephalhematoma, subgaleal hematoma,
subdural hemorrhage, parenchymal hemorrhage, intracranial
hemorrhage, retinal hemorrhage.
Maternal Injuries: Vaginal, cervical, uterine, bladder, rectal
tissue tears.
®
®
II M-Style
Mushroom
®
Cup
PREREQUISITES FOR VACUUM-ASSISTED VAGINAL DELIVERY
Term infant
Ruptured membranes
Empty bladder and nondistended rectum
Engaged vertex position
Complete cervical dilation and effacement
Willingness to abandon procedure and plan for alternate
delivery route
These instructions are intended as general guidelines.
Practitioners should refer to current institutional and recognized
guidelines that address vacuum-assisted delivery.
VACUUM-ASSISTED VAGINAL DELIVERY GUIDELINES
1. Check the integrity of the Mityvac Mystic II Obstetrical
Vacuum-Assist Delivery System by pressing the cup to the palm
of your gloved hand and applying vacuum. The gauge needle
should remain steady. (NOTE: The gauge needle should rest in
the "ZERO" box when vacuum is not applied.)
CAUTION: DO NOT USE DEVICE IF GAUGE NEEDLE DOES
NOT INDICATE ZERO IN THE ABSENCE OF A VACUUM.
2. Carefully examine fetal presentation and position prior to
positioning the cup over the flexion point.
CAUTION: DO NOT USE PRODUCT IF FLEXION POINT IS
NOT ACCESSIBLE.
3. Wipe the scalp as clean as possible.
4. To insert the M-Style Mushroom Cup, fold the cup against the
stem at a 90° angle (see Figure 1).
5. With the stem in the bent
position, grasp the mushroom
shaped portion of the cup with
the fingers. The stem should
be parallel to palm and wrist.
Separate the labia with the
other hand. Continuing with the
stem in the bent position, press
the cup portion downward and
inward over the fourchette to
make contact with the fetal
scalp. Apply the center of
the cup over the flexion
point (see Figure 2).
The flexion point can be
located by identifying
the posterior fontanelle
and then moving the
finger anteriorly along
the sagittal suture
approximately 1.2 in.
(3 cm). The sagittal suture
should pass under the middle of the cup.
CAUTION: NEVER APPLY THE CUP TO ANY PORTION
OF THE INFANT'S FACE.
6. Gently move a finger around the rim of the cup to remove
intrusion of extraneous tissues and to ensure proper
cup placement.
Figure 1 - M-Style Mushroom Cup
Posterior
Fontanelle
3 cm
Flexion Point
6 cm
Sagittal Suture
Anterior Fontanelle
Figure 2 - Infant head
10057-DFU• Rev. A • 8/13
loading

Inhaltszusammenfassung für Mityvac 10057

  • Seite 3 „There should be a willingness to abandon attempts at vacuum RICHTLINIEN FÜR VAGINALE SAUGGLOCKENENTBINDUNG extraction if satisfactory progress is not made.“ (Bei Ausbleiben Prüfen Sie die Integrität des geburtshilflichen Mityvac Mystic II eines zufriedenstellenden Fortschritts muss die Bereitschaft Saugglockenentbindungssystems, indem Sie auf die Saugglocke auf bestehen, die Versuche zur Saugglockenentbindung aufzugeben.)
  • Seite 4 , Mystic II, M-Style und Mushroom sind eingetragene Marken Prüfen Sie die Integrität des geburtshilflichen Mityvac Mystic II von CooperSurgical, Inc., deren Tochtergesellschaften und/oder Saugglockenentbindungssystems, indem Sie auf die Saugglocke auf verbundenen Unternehmen. die Handfläche Ihrer behandschuhten Hand drücken und ein Vakuum anlegen.