Immunity Test standard
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient / burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, interruptions and
variations on power supply input
lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic fi eld
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE: UT is the a.c. mains voltage prior to application of the test level.
02392_DE_180x140_GM_V1.indd 21
02392_DE_180x140_GM_V1.indd 21
IEC 60601-1-2 Test Level
±8 kV contact
±2 kV, ±4 kV,±8 kV, ±5 kV air
±2 kV, for power supply lines
±1 kV
100 kHz repetition frequency
±0.5 kV, ±1 kV diff erential lines
±0.5 kV, ±1 kV , ±2 kV common mode
0 % UT; 0.5 cycle. At 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT; 25 / 30
cycles; Single phase: at 0°.
0 % UT; 250 / 300 cycle
30 A/m
50 Hz / 60 Hz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Compliance Level
±8 kV contact
±2 kV, ±4 kV,±8 kV, ±5 kV air
±2 kV, for power supply lines
Not applicable
100 kHz repetition frequency
±0.5 kV, ±1 kV diff erential lines
Not Applicable
0 % UT; 0.5 cycle. At 0°, 45°, 90°,
135°, 180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT; 25 / 30
cycles; Single phase: at 0°.
0 % UT; 250 / 300 cycle
30 A/m
50 Hz / 60 Hz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur radio bands
between 0.15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz