Pipetting:
1.
Aspirate
Carefully hold the pipette and securely insert it into the
pipette holder in the nose cone. Select the pipetting
speed by rotating the speed control wheel. Press the as-
pirating button. The flow rate is regulated by the pressure
applied to the aspiration button.
2.
Dispense
In "Gravity mode", press the dispensing button lightly
(4 ~ 5 mm). The liquid in the pipette is dispensed by
gravity only.
3.
In "Blow-Out mode", press the dispensing button (> 5
mm). The flow rate is regulated by the pressure applied to
the dispensing button.
Note: Speed control should depend on the pipette used. Plea-
se do not use the controller without the protective membrane
filter. Replace the filter if the liquid pass it.
4.
Self-inverted Design
The pipette controller can be self-inverted on the bench
without a holder.
Maintenance
Cleaning
Before cleaning, disconnect the unit from the charging adaptor.
The controller can be cleaned with most common laboratory
disinfectants, such as 60% Isopropanol, 70% ethanol or mild
detergent. The PTFE filter, polypropylene nose cone and silicone
pipette holder can be autoclaved (121°C, 1.2 kg/cm
Filter/Pipette Holder Replacement
The hydrophobic PTFE filter and pipette holder will require
occasional replacement, particularly when aspirating speed is
decreased. Rotate the nose cone clockwise and remove it. You
can squeeze the top of the pipette holder and then remove the
pipete holder and filter together. Replace the wet, dirty filter/
worn pipette holder with a new one. Reassemble the device.
Troubleshooting:
To ensure of the product's quality and performance, the pipette
controller is designed to be robust and long-lasting. If the
problems cannot be resolved with the solutions below, please
contact your nearest Greiner Bio-One office or authorized
distributor.
Symptom
Possible Cause
Reduced suction
Filter blocked,
efficiency.
dirty or wet.
Low battery.
Pipette drips.
Pipette not se-
curely inserted in
pipette holder.
Pipette holder/
filter inserted
incorrectly.
Pipette holder or
pipette damaged.
No function even
Battery is aged.
though the battery
is fully charged.
Pipette loose.
Pipette holder
damaged.
Storage
The pipette controller should be stored in a dry place. The
allowable storage temperature: +10°C to +35°C or 50°F to 95°F.
Do not store the controller with a filled pipette.
Warranty and Service:
The Sapphire Maxipette are covered by a warranty for two years
against defects in materials and workmanship. This period
begins from the date of purchase, and within this period all
defective parts will be replaced at no charge by the manufactu-
rer. The warranty does not cover defects caused by excessive
wear and tear or damage due to shipping, accident, abuse,
misuse, problems with electrical power, or usage not in accor-
2
for 20 min).
dance with product instructions, or if other than original spare
parts supplied by the manufacturer have been used. In the event
that service or technical support is required, please contact your
nearest Greiner Bio-One office or autorized distributor. When
returning the Sapphire MaxiPette, ensure that the device is fully
decontaminated and does not present any kind of health risk to
our service personnel.
Disposal:
In the case the product is to be disposed of, the relevant legal
regulations are to be observed. The disposal of electrical devices
is regulated within the European Community by national regula-
tions based on ED directive 2002/96/ EC on waste electrical and
Solution
electronic equipment (WEEE). As disposal regulations within EU
Clean or replace
may vary from country to country, please contact your supplier
filter.
if neccessary.
Recharge battery.
Replacement Equipment:
Insert pipette
further.
Item No.
Description
847900-18
Replacement Filter
Reinsert the
pipette holder and
847903-18
Nose cone
filter carefully.
847907-18
Pipette Holder
Replace pipette
847910-18
Lithium Battery
holder or pipette.
Replace battery.
Replace pipette
Greiner Bio-One Contact details:
holder.
Germany
Greiner Bio-One GmbH
Phone +49 7022 948-0
Fax +49 7022 948-514
E-Mail info@de.gbo.com
Austria
Greiner Bio-One GmbH
Phone +43 7583 6791-0
Fax +43 7583 6318
E-Mail office@at.gbo.com
Belgium
Greiner Bio-One BVBA/ SPRL
Phone +32 2461 0910
Fax +32 2461 0905
E-Mail info@be.gbo.com
Brazil
Greiner Bio-One Brasil
Phone +55 19 3468-9600
Fax +55 19 3468-9621
E-Mail office@br.gbo.com
China
Greiner Bio-One Suns Co., Ltd.
Phone +86 10 83 55 19 91
Fax +86 10 63 56 69 00
E-Mail office@cn.gbo.com
France
Greiner Bio-One SAS
Phone +33 1 69 86 25 25
Fax +33 1 69 86 25 35
E-Mail infos@fr.gbo.com
Hungary
Greiner Bio-One Hungary Kft.
Phone +36 96 213 088
Fax +36 96 213 198
E-Mail office@hu.gbo.com
Japan
Greiner Bio-One Co. Ltd.
Phone +81 3350 58875
Fax +81 3505 8974
E-Mail info@jp.gbo.com
Netherlands
Greiner Bio-One B.V.
Phone +31 172 4209 00
Fax +31 172 4438 01
E-Mail info@nl.gbo.com
United Kingdom
Greiner Bio-One Ltd.
Phone +44 1453 8252 55
Fax +44 1453 8262 66
E-Mail info@uk.gbo.com
USA
Greiner Bio-One North America Inc.
Phone +1 704 261-7800
Fax +1 704 261-7899
E-Mail info@us.gbo.com
Revision: July 2018