Troubleshooting

The power supply has built-in error messages to help troubleshoot problems with the
electrical circuit. The error messages, and possible corrective actions, are listed below:
Error code
Explanation
E1
No Load Detected
The voltage leads to the electrophore-
sis cell are not attached.
The current load for the separation
is below .01 A, the lower limit of the
power supply.
E3
Change in Load Resistance
Open circuit was detected at some
point after a successful start.
Buffer is leaking from the cells.
There is a loose electrical connection
in the circuit.
E4
Power Failure during a continuous run
(no timer being used)
E5
Power Failure during a timed run with
PFD option disabled.
Run was not completed.
E6
Power Failure during a timed run with
PFD option enabled.
Despite the power outage
the run was completed successfully.
E7
Short Circuit
Load current exceeded 2 A.
*Press any button, except the START/STOP button, to clear the Error Message.
Other Problems:  
No display and/or fan not operating: Check that the power supply is plugged into a working
receptacle. Check that the main power switch is on. Check fuses.
Possible solutions
Check all connections.
Verify that buffer volumes are sufficient to
contact all electrode wires and that buffer is
contacting both surfaces of the gel.
A constant volt run with low volt setting may
result in a current below the .01 A limit.
Press any button* to clear the error code.
Check and correct all potential resistance
problems.
Check buffer volumes.
Check all voltage leads.
Check the fit of power supply adapters, if used.
The output current may have fallen below the
.01 A limit of the power supply.
Press any button* to clear the error code.
Restart the run, if desired.
Press any button* to clear the error code.
Restart the run, if desired.
Press any button* to clear the error code.
See page 12 for information on the PFD option.
Press any button* to clear the error code.
See page 12 for information on the PFD option.
Check for any short circuit or evidence of
electrical arcing.
Buffer concentration too high.