Herunterladen Inhalt Inhalt Diese Seite drucken

Cleaning And Disinfection - FLAEM NUOVA SpA NEBULFLAEM F2000 Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 13
GB

CLEANING AND DISINFECTION

Accessories
Carefully wash all parts used with the medicament under running water with mild (non abrasive) detergents or in the
dishwasher.
All accessories (except for the connecting tube) can be disinfected or sterilised choosing one of these methods:
a) by boiling in a bain-marie (for max 5 min.); this method cannot be used for the mask
b) by soaking in a solution of water and vinegar (3/4 water, 1/4 vinegar). Rinse thoroughly with drinking water.
• The RAPIDFLAEM 4 nebulizer is autoclavable (see 'Technical features')
• Should the valves in the RAPIDFLAEM 4 nebulizers become blocked, disassemble the individual parts, clean and
disinfect them as described above.
Unit
Wipe the case only with a damp cloth using non abrasive detergents.
The unit is equipped with an air filter (5) which should be replaced when dirty. Use only Flaem's original accessories.
IMPORTANT!!
Microbial contamination
A personal use of the nebulizer and of the accessories is recommended in order to avoid any risks of infection.
Interactions
The materials used that come into contact with drugs have been tested with a wide range of drugs. However, due to
the variety and continuous evolution of the drugs, the possibility of a chemical interaction cannot be excluded. Once
opened, the drug should be used as soon as possible, and prolonged contact between drug and nebulizer should be
avoided.
Disposal: to safeguard the environment, when the appliance has reached the end of its life take it to an authorized
electric appliance disposal center where the raw materials can be recycled.
UNIT TECHNICAL FEATURES
Mod. NEBULFLAEM F2000
~
Voltage :
230V
50Hz 210VA
Fuse :
T 2A - 250V
In compliance
Dir. 93/42/EEC
with:
(EN 60 601-1/A2)
0051
Approvals:
Operating pressure (with nebulizer):
Max pressure:
Compressor air output:
Dimensions:
Weight:
Sound level (at 1 m):
Continuous use
Operating requirements: temperature min. 10°C; max 40°C
Storage requirements:
Operating/storage
atmospheric pressure:
Electromagnetic compatibility
This aerosol therapy unit is designed to meet the current requirements for electromagnetic compatibility. However if you
suspect that the unit is interfering with the operation of your TV, radio or other electrical appliance, try relocating the unit
until the interference ceases. Alternatively try plugging the unit into a different power socket.
Warranty: The unit is under warranty from the dealer that sold you the unit.
The Manufacturer, the Retailer and the Importer are responsible for safety, reliability and performance of the unit only if: •
the unit is used in accordance with the instructions for use; • the electrical system of the facility where the unit is used adheres
to IEC standards; • assembly, additions, recalibrations, modifications or repairs are carried out by authorized personnel.
The manufacturer reserves the right to introduce technical changes.
~
115V
60Hz 190VA
T 3A - 250V
1.1 bar approx.
3.5 bar approx.
14 LPM approx.
20x27x10.5 cm
2.300 Kg
55 dB (A)
air humidity min. 10%; max 95% RH
temperature min. -25°C; max 70°C
air humidity min. 10%; max 95% RH
min. 690 hPa; max. 1060 hPa P
~
220V
60Hz 200VA
100V
T 2A - 250V
SYMBOLS
LATEX
6
~
50/60Hz 200VA
T 5A - 250V
Class II unit
Type B unit
Warning: follow the
"Instructions for use"
Alternating current
Switch OFF
Switch ON
Never use the unit while
having a bath or shower
Latex Free

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis