BUFFALO FILTER
ViroSafe
Fluid Trap
®
Instructions For Use
Designed for use with Buffalo Filter ViroSafe
except VS12D01 and VS12D02
PlumeSafe
Turbo
®
CAUTION: Ensure fluid trap is fully seated into the filter prior to capturing fluid.
Filters
®
1. Remove the fluid trap from its packaging.
2. Remove rear cap (side without the fluid
measurements) and place on appropriate
cap holder on top of fluid trap.
3. Open the 22 mm (7/8 in) port on the ViroSafe
and insert the fluid trap, as shown, into the filter.
Make certain the fluid trap is fully seated into the filter.
4. Remove the appropriate cap on the front of the fluid
trap based on the size of the tubing that will be used
in the surgical application. Store the cap on the
appropriate cap holder on top of fluid trap.
en
filter
®
loading

Inhaltszusammenfassung für Buffalo Filter FUMOVAC

  • Seite 3 BUFFALO FILTER FUMOVAC Filter Flüssigkeitsfalle Gebrauchsanweisung 1. Flüssigkeitsabscheider aus der Verpackung nehmen. 2. Die rückseitige Kappe (die Seite ohne Flüssigkeitsmessungen) abnehmen und auf entsprechenden Kappenhalter oben auf dem Flüssigkeitsabscheider stecken. 3. Den 22 mm (7/8 Zoll) Anschluss am Filter öffnen und den Flüssigkeitsabscheider (wie dargestellt) in den Filter einführen.
  • Seite 4 Nur zum Einmalgebrauch. Wiederverwendung, Wiederaufbereitung oder erneutes Sterilisieren kann zum Versagen des Geräts oder zur Kreuzkontamination führen. unter bf.customerfeedback@filtrationgroup.com an Buffalo Filter oder an Ihren örtlichen Händler. Zusätzlich sollten europäische Kunden auch an europe@emergogroup.com und an die zuständige Behörde in dem jeweiligen Mitgliedsstaat berichten.