5. Calibration
If you change the measuring range between fresh and sea water, the unit has to be recalibrated.
For fresh water, a calibration solution of 250 µS and for salt water a solution of 35 mS is
recommended. Immerse the electrode in a calibration liquid of the corresponding range and the
exact temperature. To adjust the display, turn the calibration knob (6 or 7) of the corresponding
range.
For fresh water, a calibration second fluid with 500 µS/cm, for sea water with 50 mS/cm is
necessary. Turn a small screw driver through the opening at the bottom of the unit until the display
shows the right value.
If the unit shows a wrong value, it has to be recalibrated.
6. Electrode
The conductivity electrode has a plactic shaft. The contact pins are made from carbon and are
corosion safe, also in sea water in long term use. The electrode is supplied with a thermosensor for
automatic temperature compensation. The electrode is maintenance-free.
Technical Data
Display:
Measuring range:
Resolution:
Accuracy:
Max. power on relay:
Working temperature:
Temperature compensation:
Humidity:
Control range:
Control accuracy:
Power supply:
Dimensions:
Weight:
7. Warranty
Should any defect in material or workmanship be found within 24 months of the date of purchase
AB Aqua Medic GmbH undertakes to repair or, at our option, replace the defective part free of
charge – always provided the product has been installed correctly, is used for the purpose that was
intended by us, is used in accordance with the operating instructions and is returned to us carriage
paid. The warranty term is not applicable on the all consumable products.
Proof of Purchase is required by presentation of an original invoice or receipt indicating the dealer's
name, the model number and date of purchase, or a Guarantee Card if appropriate. This warranty
may not apply if any model or production number has been altered, deleted or removed,
unauthorised persons or organisations have executed repairs, modifications or alterations, or
damage is caused by accident, misuse or neglect.
We regret we are unable to accept any liability for any consequential loss.
Please note that the product is not defective under the terms of this warranty where the product,
or any of its component parts, was not originally designed and / or manufactured for the market in
which it is used. These statements do not affect your statutory rights as a customer.
If your AB Aqua Medic GmbH product does not appear to be working correctly or appears to be
defective please contact your dealer in the first instance. If you have any questions your dealer
cannot answer please contact us.
Our policy is one of continual technical improvement and we reserve the right to modify and adjust
the specification of our products without prior notification.
AB Aqua Medic GmbH - Gewerbepark 24 - 49143 Bissendorf/Germany
6
0.5 " LED, 3-1/2 digits
0 – 1,999 µS
0 - 199,9 mS
1 µS ( 0 – 1,999 µS)
0,1 mS ( 0 - 199,9 mS)
+/- 1% (+ 2 digits)
5 Amp. at 240 V (1,000 W)
0 - 50 °C
0 - 50 °, automatic
below 80%
0 - 199 mS, digitally adjustable
+/- 5 µS, +/- 0,5 mS(+ 1 digit)
220 V, 50 Hz
150 x 85 x 40 mm
550 g
- Technical changes reserved – 07/2015