Herunterladen Diese Seite drucken

Weidmuller DI350 Bedienungsanleitung Seite 10

3 1/2-stellige digital-anzeige mit hilfsstromversorgung

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

•Displayscaling(calibration)
General
To set the display scaling you select the decimal point link position and
adjust the ZERO (Z) and SPAN (S) control to give the required units.
The ZERO adjusts the 'zero' point reading (adjustment range ±100
counts). The SPAN adjusts the 'full scale' reading (adjustment range 0
to 100 counts).
Procedure
1. If the decimal point needs to be changed, remove the back plate by
gently prising apart the four clips that hold it in place, place the link
(on the display board) as shown (to right) and reassemble.
. Inject a 'zero' (normally 4 mA) signal and adjust the Z control to give
the required display value.
. Set the input source to 'full scale' (normally 0 mA) and adjust the S
control to give the required full scale display.
4. Recheck the zero and repeat steps  &  if required.
Dimensions
92.0 mm (+0.8 / -0.0 mm)
CUTOUT
10
18.0 mm
GASKET
Donotovertightenthemountingscrews
PANEL THICKNESS Max 18.0 mm

Werbung

loading