6) SPECIFICATIONS

General Specifications
Display
Polarity
Update Rate
Operating Temperature
Relative Humidity
Altitude
Pollution degree
Storage Temperature
Temperature Coefficient
Sensing
Safety
Terminals (to COM) measurement category:
Overload Protections :
Transient protection
Power Supply
Power Consumption
Low Battery
E.M.C.
Elma 805/807
3-3/4 digits 4000 counts LCD display
Automatic
3 per second nominal
0C to 40C
Maximum relative humidity 80% for temperature up
to 31C decreasing linearly to 50% relative humidity
at 40C
Operating below 2000m
2
-20C to 60C, < 80% R.H. (with battery removed)
Nominal 0.15 x (specified accuracy)/ C @(0C -18C
or 28C -40C), or otherwise specified
Average sensing for Models 805s. True RMS for
Model 807s
The meter protection rating, against the users, is
double insulation per IEC/UL/EN61010-1 Ed. 3.0,
IEC/EN61010-2-030 Ed. 1.0, IEC/EN61010-2-033
Ed.
1.0,
IEC/UL/EN61010-031
CAN/CSA-C22.2 No. 61010-1-12 Ed. 3.0 to Category
II 1000 Volts, CAT III 600Volts and CAT IV 300Volts
AC & DC.
V / A / mAA : Category II 1000V, CAT III 600V and
CAT IV 300V AC & DC.
A & mA : 0.4A/1000V DC/AC rms, IR 30kA @
1000V DC/AC rms
A : 11A/1000V DC/AC rms, IR 20kA @ 1000V
DC/AC rms
V : 1100V DC/AC rms
Hz, Ohm & others : 1000V DC/AC rms
6kV (1.2/50s surge)
1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
3.2 mA typical
Below approx. 2.4V
Meets
EN61326(1997,
2(1995), and EN61000-4-3(1996)
In an RF field of 3V/m: Capacitance function is not
specified
AC 4.000V range: Total Accuracy = Specified
Accuracy + 700 digits