SPECIFICATIONS
Method -
Dual Beam NDIR (Non-Dispersive-Infrared)
Display - LCD Independent CO
Sample Method - Diffusion or flow through (50 ~200 ml/min).
CO
rH & Temperature Specification:
2,
CO
Specification:
2
Measurement Range
Display Resolution
Accuracy
Repeatability
Temperature Dependence
Response Time
Warm-Up Time
Zone LED Display
Temperature Specification:
Temperature Range
Display Resolution
Display Options
Accuracy
Response Time
rH Specification:
Measurement Range
Display Resolution
Accuracy:
Response time:
Operating Conditions:
Operating Temperature
Humidity Range
Storage Conditions:
Storage Temperature
Power Supply:
Power Supply
DC Output voltage
AC Input voltage & Frequency
Dimensions
Error Codes & Troubleshooting Guide
Error Code
The ambient temperature has exceeded the operating
「
」
temperature range 0°C to 50°C (32°F to 122°F)
「
」
EEPROM System Problem
, rH and temperature readings
2
0-3,000 ppm display
1ppm at 0~1,000ppm; 5ppm at 1,001~2,000ppm; 10ppm at 2,001~3,000ppm
Under 25°C: ±80 ppm or ±5% of reading (≦2000ppm); ±7% reading (>2000ppm)
±20 ppm @ 400ppm
Typ. ±0.2% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C
About 2min for 63% of step change
2 min. at 22°C
Green: <800ppm; Yellow: 800~1200ppm;Red: >1200ppm
0°C to 50°C (32°F to 122°F) display
0.1°C (0.1°F)
°C/°F
±1°C( ±2°F) When the fan blows to the device directly, the accuracy of temperature is + / -1.5 degC
20-30 minutes (case must equalize with environment)
20%-90%rH
1%rH
±5%rH@23°C
<5 min for 63% of step change
0°C to 50°C (32°F to 122°F)
0 ~ 95% rH non-condensing
-20°C to 60°C( -4°F to 140°F)
USB or 5 VDC from external AC/DC adapter which is included in package (Use specified AC adapter only)
5 VDC
100 ~ 240 VAC, 0.2A 50 / 60 Hz
Description of the fault
WTW GmbH – Geschäftsbereich ebro Electronic
Peringerstr. 10 – 85055 Ingolstadt - Germany.
Tel. +49 841 95478-0
ebro@xyleminc.com
http://www.ebro.com
Suggested Actions
This error will clear when the temperature returns to the range between
0°C to 50°C (32°F to 122°F).
Please reconnect AC adapter to the RM 100.
If the "Err5, Err6" still appears, please contact the Service Department for
further assistance.