5.3.6 Sensitivity coefficient of sample gas temperature

The sensitivity coefficient of sample gas temperature at reference point shall not ex-
ceed the requirements of Table 2 of VDI 4202, Sheet 1 (September 2010). A value
c
at 70 % to 80 % of the upper limit of the certification range shall be used as refer-
t
ence point.
The sensitivity coefficient of sample gas temperature shall not exceed
0.3 (µmol/mol)/K (i.e. (0.35 mg/m³)/K).
6.2
Equipment
Not applicable here.
6.3
Testing
Performance and evaluation of the steps taken to determine the sensitivity coefficient of
sample gas temperature are in line with the requirements stipulated in DIN EN 14626 (2012).
The reader is therefore referred to section 7.1
temperature.
6.4
Evaluation
Please refer to section 7.1
6.5
Assessment
Please refer to section 7.1
Does this comply with the performance criterion? yes
6.6
Detailed presentation of test results
Not applicable in this instance.
EP_Ecotech_Serinus30_en.docx
Report on the performance testing of the Serinus 30 ambient air quality
monitoring system manufactured by Ecotech Pty Ltd measuring CO, Report
8.4.8 Sensitivity coefficient to sample gas
8.4.8 Sensitivity coefficient to sample gas temperature.
8.4.8 Sensitivity coefficient to sample gas temperature.
TÜV Rheinland Energie und Umwelt GmbH
no.: 936/21221977/D_EN
Luftreinhaltung

Kapitel