Seite 1
User Manual checks as part of our QM documentation. Water-i.d. GmbH (Germany) Manuel d’utilisation Andreas Hock, Managing Director Water-i.d. GmbH • Daimlerstr. 20 • D-76344 Eggenstein • Germany Water-i.d. is certified according to ISO 9001:2016 Manufacturing date...
Seite 2
TEST - Hyd. Peroxide (H O ) 34 - 35 Kühl und trocken lagern! OR / UR / Dilution Troubleshooting (Error) Reagents for water-analysis only! Do not eat! Keep out of reach of children! Changing cuvette / calibration Store cool and dry! Accessories App / Software Utiliser uniquement des réactifs...
Seite 3
Batterien • Batteries • Piles Einschalten • Switch on • Allumer • wechseln • change • chargement 2 sec. · ZERO Der On/Off Knopf kann auch zum Abbrechen des Countdowns während der Messung verwendet werden (nicht empfohlen) On/Off button can also be used to skip countdown during measurement (not recommended) Le bouton Marche / Arrêt peut être également utilisé...
Seite 4
Hinweise • Advices • Conseils Einschalten • Switch on • Allumer IMMER Photometer-Tabletten und NIE RAPID-Tabletten verwenden! Die Tabletten dürfen nicht Detergent berührt werden! waterproof (IP68 ) PHOTOMETER Always use PHOTOMETER ZERO grade tablets! Never use RAPID grade tablets! Do not touch reagent tablets! Es ist wichtig, das Gerät nach jeder Messung zu reini- Toujours utiliser des pastilles gen, um sämtliche Reagenzienrückstände zu...
Seite 5
ZERO Nur 1 x pro Testreihe • Only 1 time per test batch 3 x = • Une seule fois par lot de test waterproof (IP68 ) ZERO muss nur einmal pro Testreihe durchgeführt ZERO werden. Sobald erfolgt, können alle folgenden Messungen (z.B.
Seite 6
Nach / After / Après ZERO completely Phenol Red dissolved Photometer Range: 6.50 - 8.40 pH TEST! TEST! Reagent: Phenol Red Photometer residue ZERO ZERO 3 x = waterproof (IP68 ) ZERO 10 ml 15 sec. 7.10 waterproof TEST! (IP68 ) 10 ml ·...
Seite 7
Nach / After / Après ZERO Freies Chlor • Free Chlorine Chlor • Chlore libre N° 1 Chlorine Photometer completely Chlore dissolved Range: 0.00 - 6.00 mg/l (ppm) Reagents: DPD N° 1 Photometer TEST! TEST! Reagents: DPD N° 3 Photometer ZERO ZERO residue...
Seite 8
ZERO Total Chlorine is measured directly after free Chlorine without emptying the cuvette. The DPD 3 tablet is added to the sample water which already contains the DPD 1 tablet (dissolved). Combined Chlorine is calculated as Total Chlorine minus free Chlorine.
Seite 9
Nach / After / Après ZERO Cyanursäure • Cyanuric Acid Cyanursäure • Acide cyanurique Cyanuric Acid CYA-Test completely Photometer dissolved Acide cyanurique Range: 0 - 160 mg/l (ppm) Reagent: CYA-Test Photometer TEST! TEST! ZERO ZERO residue 3 x = waterproof (IP68 ) ZERO 10 ml...
Seite 10
Nach / After / Après ZERO Alkalinität • Alkalinity • Alcalinité Alkalinität Alkalinity Alkalinity-M Photometer Alcalinité completely dissolved Range: 0 - 300 mg/l (ppm) CaCO Reagent: Alkalinity-M Photometer TEST! TEST! ZERO ZERO 3 x = residue waterproof (IP68 ) ZERO 10 ml 15 sec.
Seite 11
Nach / After / Après ZERO Aktivsauerstoff • Active Oxygen Aktivsauerstoff (MPS) • Oxygène actif N° 4 Active Oxygen (MPS) Photometer* Oxygène actif (MPS) completely dissolved Range: 0.3 - 30.0 mg/l (ppm) TEST! Reagent: DPD N° 4 Photometer* TEST! ZERO ZERO *not part of standard equipment residue...
Seite 12
Verfahren "B" folgen. Reagent: DPD N° 1 Photometer Glycine* Only if your water sample does contain Chlorine next to Chlorine Dioxide (both disinfectants used), *not part of standard equipment the following procedure „A“ needs to be followed and Glycine reagent needs to be used. Otherwise...
Seite 13
Nach / After / Après ZERO Nach / After / Après ZERO Chlordioxid • Chlorine Dioxide Chlordioxid • Chlorine Dioxide • Dioxyde de Chlore • Dioxyde de Chlore Mit Chlor / With Chlorine / Avec du Chlore Ohne Chlor / Without Chlorine / Sans Chlore DPD N°...
Seite 14
Reagent: DPD N° 1 Photometer Reagent: Glycine Only if your water sample does contain Chlorine next to Bromine (both disinfectants used), the *not part of standard equipment following procedure „A“ needs to be followed and Glycine* reagent needs to be used. Otherwise (only 3 x = Bromine present), please follow procedure „B“...
Seite 15
Nach / After / Après ZERO Nach / After / Après ZERO Brom • Bromine • Brome Brom • Bromine • Brome Mit Chlor / With Chlorine / Avec du Chlore Ohne Chlor / Without Chlorine / Sans Chlore DPD N° 1 Glycine* Photometer completely...
Seite 16
Reagent: DPD N° 3 Photometer Reagent: Glycine Only if your water sample does contain Chlorine next to Ozone (both disinfectants used), the *not part of standard equipment following procedure „A“ needs to be followed and Glycine* reagent needs to be used. Otherwise (only 3 x = Ozone present), please follow procedure „B“...
Seite 17
Nach / After / Après ZERO Nach / After / Après ZERO Ozon • Ozone Ozon • Ozone Mit Chlor / With Chlorine / Avec du Chlore Ohne Chlor / Without Chlorine / Sans Chlore Photometer Glycine* DPD N° / DPD N° completely completely dissolved...
Seite 18
Nach / After / Après ZERO Wasserstoffperoxid • Hydrogen Peroxide • Peroxyde d'Hydrogène Wasserstoffperoxid Hyd. Peroxide LR Hydrogen Peroxide Photometer* Peroxyde d'Hydrogène completely dissolved Range: 0.3 - 2.9 mg/l (ppm) TEST! Reagent: Hyd. Peroxide LR Photometer* TEST! ZERO ZERO *not part of standard equipment residue 3 x = waterproof...
Seite 19
Use syringe to take only 5ml (or 1ml) sample • (Trop sombre) Nettoyer la chambre de mesure ou diluer l'échantillon water plus 5ml (9ml) distilled water. test again and multiply results times 2 (times 10). DIlution does not • (zu hell) Lichtschutzdeckel bei der Messung nicht Err03: work with „pH“...
Seite 20
Küvettenwechsel • Changing the cuvette Zubehör • Accessories • Accessoires • Changer la cuvette Reagenzien • Reagents • Réactifs POL01-Nf 20/20/10/10/10 90° Phenol Red / DPD N° 1 / DPD N° 3 / CYA-Test / Alkalinity-M Photometer TbsPph100 100 x Phenol Red Photometer TbsPD1100 100 x DPD N°...
Seite 22
According to EC Directive 2002/96/EC, electronic devices must not be disposed of in normal domestic waste. The manufacturer of this device, Water-i.d. GmbH, Daimlerstr. 20, D-76344 Eggenstein will dispose of your 9in1 Multitest Photometer free of charge (not including costs of sending the device to us). Send your 9in1...
Seite 23
- Connect the equipment into an outlet on circuit different from that to which the receiver is connected. Changes or modications not expressly approved by Water-i.d. - Consult the dealer or an experienced radio/TV technician for GmbH could void the user‘s authority to operate the equipment.