Track-It
TM
SAFEGUARDS AND PRECAUTIONS:
Read and follow all instructions in this instruction sheet carefully, and retain this sheet for future reference.
Do not use this instrument in any manner inconsistent with these operating instructions or under any
conditions that exceed the environmental specifications stated.
This instrument is not user serviceable. For technical assistance, contact the sales organization from which
you purchased the product.
DESCRIPTION:
The Pressure/Vacuum Temperature Loggers are sealed (IP67) Data Loggers used for measuring air or
liquid pressure and ambient temperature. The units are self contained in a small 316L stainless steel
enclosure with a threaded ¼" NPT port to connect to a pressure source. The Loggers can operate
unattended for months or years. Recording parameters can be set in the Track-It
units can be set to record on instantly, at preset times, or on alarm conditions. Up to 64,000 readings can
be saved in non-volatile memory at intervals from 2 seconds to 24 hours. Internally there is a USB port
for direct connection to a laptop or PC, and a button and LED for user interaction as described below.
SPECIFICATIONS:
General
Record:
Sample Rates: User configured 1 every 2 seconds up to 1 every 24 hours
No. of Samples: 32,000 or 64,000 (depends on setup)
Record Trigger:
Two independent triggers. Multiple trigger modes: Instantaneous, button control, on alarm, time and date
(start and stop), day of week
Record Mode:
Fill to end of memory, cyclic, number of samples, time duration
Alarms:
2 user programmable alarms. High or Low
Communication: Direct USB connection
Software:
Track-It™ Software—Program device, view data (historic or real time), export to Excel™.
Battery:
Lithium 1/2AA 1.2Ah
Life: Up to 5 years typical @ 1 minute sample rate
Enclosure:
Material: 316L Stainless Steel
Dimensions: 1.95" (5cm) diameter x 2.19" (5.6cm) excluding port thread
Measurement
Ranges:
Pressure: 0-35, 0-150, 0-350, 0-550, 0-2000, 0-5800 PSI
Vacuum: 760-0 Torr and 760-380 Torr
Accuracy:
Resolution:
Repeatability:
Overpressure Rating: 1.5 times max
Port connections:
Temperature Range: -20 to +85°C /-4 to 185°F
Accuracy:
± 0.5°C 0 to 50°C / 32 to 122°F
± 2°C -20 to 85°C / -4 to 185°F
Resolution:
0.06°C / 0.11°F
INSTALLATION:
The logger screws into a ¼" NPT port. Use suitable tape or thread sealant/dope on the threads before inserting the logger into
the port. Use an adjustable wrench on the nut to tighten the device into the port and test for leaks.
Pressure/Vacuum Temperature Logger
± 0.5%
0.01%
± 0.05%
1/4" NPT Male
PC Software. The