The system comes with gown clips to relieve the weight of the cable from the headband.
One gown clip is attached below the shoulder.
Once the battery is depleted into the last shown "block" on the power gauge, the unit may
beep once to indicate low battery levels. The unit may beep twice to indicate further
reductions in power before shut-down.
4. MAINTENANCE
Storing the headlight system in a safe place will prolong the life of the unit.
The following guidelines will help in sustaining the headlight system useful life:
Keep the optical faces of the LED Headlight lens from touching hard surfaces,
which may cause scratches on the surface. Scratches on the surface will
diminish the light output. Using the lens cap during transport will help protect the
lens.
5. CLEANING
The headlight module and headband can be wiped down with commercially available
cleansers commonly used for disinfection of electronic equipment in hospitals such as ethyl
or isopropyl alcohols, disinfecting sprays containing quaternary ammonium compounds, or
hydrogen peroxide. The lens should be cleaned only with lens tissue, available in any camera
store; follow the directions on the package.
Do not use strongly caustic or acidic cleansers such as "Clorox" hypochlorite bleach,
ammonia, muriatic acid or similar products. Do not use acetone, methyl ethyl ketone or
halogenated / chlorinated hydrocarbon solvents or cleansers containing any of these
restricted compounds.
Apply cleaning agents by light spray or dampened towels. Do not pour liquids onto the
device. Do not allow liquids to enter the device seams or ventilation openings.
Follow all applicable bloodborne pathogen procedures as required by OSHA and/or your
hospital, when cleaning and disinfecting the product.
Do not autoclave the LED Headlight module, headband, or battery.
6. DESKTOP SMART CHARGER AND CALIBRATOR
Please refer to the separate battery and charger IFUs for detailed instructions on how to
safely operate these devices.
7.
REPLACEMENT PARTS
For replacement parts, additional or optional equipment :
Please call Customer Service @ 877 814 2237
®