Charging the Battery Pack
The battery pack charges automatically when the
oscilloscope is connected to line power. For a
more efficient charge, use the optional external
charger (TDS3CHG).
When the battery pack is in the oscilloscope, the
gauge icon (
) on the display indicates the
amount of charge.
S
To avoid automatic shutdown of the battery
pack, run the oscilloscope on the battery pack
only when the gauge indicates sufficient
charge.
S
To ensure the accuracy of the gauge, occasion-
ally allow the oscilloscope to run on the bat-
tery pack until automatic shutdown occurs.
NOTE.
The TDS3CHG external charger may
increase the available charge and reset the
gauge to a higher level. Subsequent charges
using the oscilloscope may indicate less than
a full charge.
Configuration
Typical charging time
Battery pack charging in
18 hours
oscilloscope with
oscilloscope turned on or off
Battery pack charging with
4 hours
TDS3CHG external charger
To avoid damage to the battery
CAUTION.
pack, use only the oscilloscope or the
TDS3CHG battery charger to charge the
battery pack. Do not connect any other
voltage source to the battery pack.
NOTE.
Charge the battery pack completely
before using it for the first time or after
prolonged storage. Several charge/discharge
cycles may be required before the battery
pack can charge to its full capacity.
NiCad batteries appear to lose capacity if not
allowed to occasionally discharge completely.
Every few months, run your oscilloscope until it
shuts down and then recharge the battery pack
completely to keep your battery pack at peak
performance.
To discharge the battery pack completely, continue
to run the oscilloscope on the battery pack until
automatic shutdown occurs. The display may turn
white a few minutes before the automatic
shutdown to allow the oscilloscope to completely
discharge the battery pack.
NOTE.
and to prevent shutdown, do not operate or
charge the battery pack at high
temperatures. For best results, allow the
battery pack to cool to room temperature
before using or charging the battery pack.
The battery pack stops charging under the
following conditions:
S
The battery pack is fully charged
S
The battery pack temperature exceeds +45_ C
S
The charge time exceeds 4.5 hours when
charging with the TDS3CHG
S
The charge time exceeds 24 hours when
charging in the TDS3000 series oscilloscope.
The battery pack shuts down and disconnects
itself from the TDS3000 series oscilloscope under
the following conditions:
S
The battery pack is fully discharged
S
There is a fault current in excess of 7A
S
The temperature inside the battery pack
compartment exceeds 70_ C
Replaceable Parts
Grounding wire
Specifications
Input from
TDS3CHG
Output
Operating
Temperature
Storage
Optional
Temperature
TDS3CHG
Temperature
Coefficient
Maximum Altitude 2000 m
To prolong the life of the battery pack
174-3997-00
30 W, 18 VDC, 1.8 A (current limited)
12 VDC, 4400 mAh
5 C to 50 C , discharging
0 C to 45 C , charging
–30 C to +50 C, short term
–20 C to +35 C, long term (greater
than 3 months
0.2 x (specified accuracy) / C
(<18 C or >28 C)
Certifications and compliances
Australia/New Zealand Declaration of Conformity – EMC
Complies with EMC provision of Radiocommunications Act
per the following standard(s):
AS/NZS 2064.1/2
Industrial, Scientific, and Medical
Equipment: 1992
EC Declaration of
Meets the intent of Directive
Conformity –
89/336/EEC for Electromagnetic
EMC
Compatibility when it is used with the
product(s) stated in the specifications
table. Refer to the EMC specification
published for the stated product. May
not meet the intent of the directive if
used with other products.
FCC Compliance
Emissions comply with FCC Code of
– EMC
Federal Regulations 47, Part 15,
Subpart B, Class A Limits.
WARRANTY SUMMARY
Tektronix warrants that the products that it
manufactures and sells will be free from defects in
materials and workmanship for a period of three
(3) months from the date of shipment from an
authorized Tektronix distributor. If a product or
CRT proves defective within the respective period,
Tektronix will provide repair or replacement as
described in the complete warranty statement.
To arrange for service or obtain a copy of the
complete warranty statement, please contact your
nearest Tektronix sales and service office.
EXCEPT AS PROVIDED IN THIS SUMMARY
OR THE APPLICABLE WARRANTY
STATEMENT, TEKTRONIX MAKES NO
WARRANTY OF ANY KIND, EXPRESS OR
IMPLIED, INCLUDING WITHOUT
LIMITATION THE IMPLIED WARRANTIES
OF MERCHANTABILITY AND FITNESS FOR
A PARTICULAR PURPOSE. IN NO EVENT
SHALL TEKTRONIX BE LIABLE FOR
INDIRECT, SPECIAL OR CONSEQUENTIAL
DAMAGES.
Copyright
Tektronix, Inc. Printed in U.S.A.
Wilsonville, OR 97070-1000

Werbung

loading