or ME SYSTEM with the requirements of Clause 7 (EMIS-
SIONS) and Clause 8 (IMMUNITY). ACCESSORIES may
be specified either generically (e.g. shielded cable, load im-
pedance) or specifically (e.g. by MANUFACTURER and
EQUIPMENT OR TYPE REFERENCE). If any: the performance
of the ME EQUIPMENT or ME SYSTEM that was determined
to be ESSENTIAL PERFORMANCE and a description of
what the OPERATOR can expect if the ESSENTIAL PERFOR-
MANCE is lost or degraded due to EM DISTURBANCES
(the defined term "ESSENTIAL PERFORMANCE" need not
be used).
Table 1:
Guidance and manufacturer's declaration –
electromagnetic emissions:
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Compliance
Group 1
Group B
Not application
Not application