3.4
Programming the Transmitter
Note
Before creating a site select English or metric units from the [Meter Facilities] menu.
Select a Meter Type
1. After power-up, use the Magnetic Wand to trigger the <ENTER> sensor and access the
2. Trigger the <Right Arrow> sensor.
3. Trigger <Up/Down Arrows> to select [2 Channel Flow]
FUH1010 IP65 NEMA 7 Interface Detector Quick Start
Operating Instructions, 06/2014, A5E02428148-AA
Store numeric data, select from option lists, etc.
Menu navigation keys move cursor.
Same as <Left> and <Right> arrows. Scrolls option lists.
ENTER Sensor
ENTER Target Symbol
top level of the Installation Menu.
Note
Select [2 Channel Flow] if measuring two different pipes and [Dual Path Flow] if sensors
are mounted on the same pipe.

3.4 Programming the Transmitter

Description
Current Selection
Top level of Installation Menu
Commissioning
23