RT-ZS60
Description
BNC to 4 mm dual banana
adapter
Two banana leads (red, black)
Probe accessories
Let the probe and oscilloscope warm up at least for 15 min prior to the test.
Self-alignment
Prior to the functional check, a self-alignment of the R&S RTO oscilloscope must
be carried out. Proceed as follows:
Test setup:
R&S RTO settings: ●
Measurement:
6.1

Initialization

The correct initialization of the probe by the oscilloscope has to be checked.
Test setup:
R&S RTO settings:
Measurement:
User Manual 1418.7342.02 ─ 02
Minimum requirements
1 signal adapter, square pin
2 mini clips
2 leads, 15 cm
R&S RTO oscilloscope
PRESET
"File" menu > "Selfalignment" > "Control" tab > "Start Alignment"
After a successful self-alignment, the label UNCAL is no longer displayed.
Probe connected to CH1 of the R&S RTO oscilloscope
"Vertical" menu > "Probe Setup" > "Ch1" tab
If initialization is correct, the following values are displayed on the
"Setup" tab of the "Probes" dialog box for the R&S RT-ZS60:
Type: active single ended
Name: R&S RT-ZS60
Probe attenuation: 10:1
Bandwidth: 6.0 GHz
Functional Check
Initialization
Recommended equipment
Pomona Model 1269
Multi-Contact ZG410
43