R&S
RT-ZS60
4.1.3
Performance with Different Connection Types
The
table 4-2
shows three types of connection between probe and DUT as well
as the associated rise times, bandwidths, input impedances and overshoots.
Table 4-2: Typical rise time, bandwidth, input impedance and overshoot with different
connection types
No
Connection
Type
1
2
3
User Manual 1418.7342.02 ─ 02
Connection
Signal
Ground
socket
socket
ground
signal
pin,
pin
pogo
signal
ground
pin, sol-
pin, sol-
der in
der in
signal
ground
adapter,
adapter,
square
square
pin
pin
Signal Integrity of the Transferred Signal
Rise
time
wide
64 ps
spacing
narrow
55 ps
spacing
short
66 ps
6.5 GHz
pins
long
70 ps
4.5 GHz
pins
wide
64 ps
5.5 GHz
spacing
narrow
52 ps
spacing
Measurement Principles
Band-
Min. input
width
impedance
|Z
|
min
150 Ω
6 GHz
110 Ω
8 GHz
155 Ω
235 Ω
120 Ω
75 Ω
9 GHz
Over-
shoot
9 %
25 %
8 %
11 %
11 %
34 %
33