Herunterladen Inhalt Inhalt Diese Seite drucken

Rohde & Schwarz R&S RT-ZS60 Bedienungsanleitung Seite 71

Inhaltsverzeichnis
®
R&S
RT-ZS60
Fig. 2-2: Offset compensation voltage and dynamic range
To set offset compensation on the front panel:
1. Press the Vertical Position knob until the "Offset" setup is shown on the dis-
play.
2. Turn the Vertical Position knob.
To automatically set the offset compensation voltage using the menu:
1. On the "Vertical" menu, select "Probe Setup".
2. Select the channel and tap "Set offset to mean".
For more details on setting the offset compensation voltage, refer to the
"User Manual" of the R&S oscilloscope.
Additionally, the probe offers automatic offset compensation by means of the
micro button. A single push on the button compensates the DC component of the
measurement signal which is particularly helpful during measurements of input
User Manual 1418.7342.02 ─ 02
Putting into Operation
Offset Compensation
20