11
Ambient Conditions
Prerequisite
Tool
KNAUER
Installation

Operating Environment

Only if the requirements for ambient conditions are met, can the
intended use be ensured.
CAUTION! 
Defect of the device due to overheating possible. 
- Set up the device in such a way that it is pro-
tected against exposure to direct sunlight. 
- Keep at least 15 cm clear at the rear and 5–
10 cm at each side for air circulation.
Sunlight: Protect the device against direct exposure to sun-
light.
AC system: Set up the device at a location not exposed to
air drafts.
Installation site: Position the device on a level and even sur-
face.
Humidity
Temperature
Recommended
working environ-
ment
Height above sea
level

Space Requirements

To protect the device from overheating, allow enough space on
each side of the device:
At least 5 cm if there is another device on one side.
At least 10 cm if other devices set up on both sides.
At least 15 cm on the rear panel for the fan.

Setup

After you have selected a suitable place for operation, remove
the device packaging. The requirements and a description can
be found in the following section.

Unpacking

Check packaging for damage caused during transportation.
Utility knife
CT 2.1 User Manual, V6810, Version 1.1
< 90 %, non-condensing
10-40 °C (50-104 °F)
17-28 °C (62.6-82.4 °F)
maximum 2000 meters