WARNING
During exchange of the proximity switch you may be exposed to ionizing radiation.
• Contact your responsible radiation safety officer for particular instructions.
• Check the source container, it must be fixed in the "OFF" position.
• For source container with pneumatic drive, make sure the pneumatic drive is unpressurized, when the cover is removed.
• Avoid any unnecessary exposure to radiation and keep unavoidable exposure to radiation as low as possible.
• Observe the instructions on radiation protection in BA01327F (FQG66).
• After the exchange switch the compressed air on and off, thereby control the function of proximity switch.
Tool list
Open terminal box
1
90°
90°
90°
90°
3 Nm
FQG6x_terminal1

Remove cover

1
Dismount proximity switch "ON"
Dismounting of proximity switch is the same for versions with pneumatic drive or manual operation.
To see more details the series of steps are described on a FQG66 with manual operation.
1
17 mm
8 Nm
FQG66_Initiator3
Endress + Hauser
17 mm
3 Nm, 8 Nm
2
2
6 mm
x 3
FQG66_cap1
2
6 mm
3 Nm
3
EIN/ON
AUS/OFF
FQG6x_terminal3
3
FQG66_cap2
3
17 mm
8 Nm
FQG66_Initiator4
3 mm
2 Nm
4
EIN/ON
17 mm
3 Nm
FQG66_Initiator1
4
FQG66_cap3
4
FQG66_Initiator5
Proximity switch
AUS/OFF
FQG6_Initiator2
FQG66_cap4
FQG66_Initiator6