6 Input C
Frequency range ..................: 100 MHz to 1.6 GHz
Resolution.............................: max. 10 Hz (gate 10s) to
min. 10 kHz (gate 0.1s)
Input sensitivity.....................: 100 MHz to 1.3 GHz => 30 mVrms
1.3 GHz to 1.6 GHz => 70 mV
Coupling ...............................: AC only
Impedance............................: 50 ohms ± 5%
Max. input voltage ................: 3 Vpp sinusoidal
7 Input C2 (D) for model MXC-2800
Frequency range...................: 1.3 GHz to 2.8 GHz
Resolution.............................: max. 100 Hz (gate 10s) to
min. 10 kHz (gate 0.1s)
Input sensitivity.....................: 1.3 GHz to 2 GHz => 30 mVrms
2 GHz to 2.8 GHz => 70 mV
Coupling ...............................: AC only
Impedance............................: 50 ohms ± 5%
Max. input voltage ................: 3 Vpp sine-shaped
Time base
Type : TCO = temperature-controlled oscillator
Frequency.............................: 10.000000 MHz
Accuracy...............................: ±1 ppm
Accuracy in relation
to mains voltage ...................: less than ±1 ppm at ±10 % mains voltage fluctuation
Temperature stability ............: ±5 ppm in the range 0°C to +50°C
Aging deviation.....................: max. ±5 ppm/year
Digital display
Display ..................................: 8-digit LED display, with the symbols M/n (mega /
nano), k/u (kilo / micro), m (milli), Sec (second), G.T.
(gate time), Hold and OF (= overflow)
Manual gate time and function setting (OF for over-
flow is displayed if the numerical range is exceeded.
Hold function ........................: For frequency or period measurements, or with the
function TOTAL, pressing the HOLD button freezes
the last measurement value until released by repres-
sing the button
10
Fixed voltage outputs of the supply unit
5 V output
Output voltage......................: 5 VDC, stabilized
Output current ......................: I
= 1 A
nom
I
= 1.3 A
max
Residual ripple......................: max. 2 mVrms
Load control .........................: 0.1% + 100 mV
Mains adjustment .................: 0.1% + 50 mV
15 V output
Output voltage of .................: 15 VDC, stabilized
Output current ......................: I
= 1 A
nom
I
= 1.3 A
max
Residual ripple......................: max. 2 mVrms
Load control .........................: 0.1% + 100 mV
Mains adjustment .................: 0.1% + 50 mV
General technical data
Environmental conditions
Working temperature range ..: 0°C to +40°C
Rel. humidity ........................: max. 85 %, non-condensing
In-store conditions................: -10°C to +50°C, under max. rel. humidity of 85%,
non-condensing
Temperature for
guaranteed accuracy............: +25°C ±5 K
Mains voltage .......................: 110/120/220/240 VAC, 50 to 60 Hz
Current consumption............: approx. 70 W (typ.) max.
Mains protection...................: 0.8 A fast-blow fuse, standard code 0.2 AF/250 V or
F0,2A/250V at 220/240V fast-blow fuse, standard
code 1 AF/250 V or F1A/250V at 110/120V
Overall dimensions (LxWxH).: approx.260 x 255 x 80 mm (excl. cables)
Weight...................................: approx. 1.8 kg (excl. cables)