Alternating voltage (V/AC) LoZ
Range
6.000 V
60.00 V
600.0 V*
Specified measurement range: 5 - 100% of the measurement range
Frequency range: 45 - 400 Hz; Overload protection: 600 V; Impedance: 400 kΩ (*max. 250 V, 3s)
The multimeter may display 10 counts if a measurement input is short-circuited
After using the LoZ feature, leave the multimeter for 1 minute before using it again
TrueRMS peak (Crest Factor (CF)) ≤3 CF to 600 V
TrueRMS peak for non-sinusoidal signals plus tolerance:
CF >1.0 - 2.0
+ 3%
CF >2.0 - 2.5
+ 5%
CF >2.5 - 3.0
+ 7%
Direct current (A/DC)
Range
600.0 µA
6000 µA
60.00mA
600.0mA
6.000A
10.00A
Overload protection: 600 V
Fuses: µA/mA = self-resetting PTC fuse 4x 160 mA, internal resistance approx. <10 Ω
10 A = High-performance FF12AH1000V ceramic fuse
≤ 6 A = continuous measurement, >6 A = max. 10 s at intervals of 15 minutes
The multimeter may display 3 counts when a measurement input is open.
72
Resolution
0.001 V
0.01 V
0.1 V
Resolution
0.1 µA
1 µA
0.01mA
0.1mA
0.001A
0.01A
Accuracy
±(2.3% + 7)
Accuracy
±(1.0% + 7)
±(1.3% + 4)
±(1.6% + 7)