Setting of Direction and Resolution Switching
The phase relationship of phases A and B and resolution for the direction of movement of
the probe spindle can be switched with the S2 switch.
(Set when shipped from the factory)
Direction switch (S2-1)
Resolution switch (S2-2)
Resolution switch (S2-3)
Resolution switch (S2-4)
Switching of Direction
Directivity is switched with S2-1 (default setting: OFF).
S2-1
ON
A:
B:
OFF
A:
B:
DG805FLM/DG805FLE
DG810FM/DG810FE
DG810FLM/DG810FLE
ON
1
2
3
Moves the probe in the direction in which
the spindle is pushed in.
B phase signal advances.
A phase signal advances.
4
(E)
7