Herunterladen Diese Seite drucken

OHAUS i-D61PW5K1S6 Bedienungsanleitung Seite 60

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

EN-58
Error 8.9
Error 8.90
Error 8.91
Error 8.92
Error 8.94
Error 8.93
Error 9.5
Battery symbol flashing
CAL E
NO.SwW
REF wWt Err
6.3
Service Information
If the troubleshooting section does not resolve your problem, contact an authorized OHAUS Service Agent. For
Service assistance in the United States, call toll-free 1-800-526-0659 between 8:00 AM and 5:00 PM Eastern
Standard Time. An OHAUS Product Service Specialist will be available to assist you. Outside the USA, please
visit our website www.ohaus.com to locate the OHAUS office nearest you.
Find Quality Products Online at:
status.
Fail to read serial number from
memory module or serial number
does not match the indicator's under
LFT ON status.
The serial number of EasyConnect
base does not match the serial
number stored inside indicator under
LFT ON status.
Fail to read data from the
TM
EasyConnect
module under LFT
ON status.
Incorrect data from EasyConnect
base under LFT ON status.
TM
The EasyConnect
base is
connected to a blank indicator under
the LFT ON status.
Calibration data not present.
Batteries are discharged.
Calibration value outside allowable
limits
Attempting to exit the menu with the
LFT setting ON and the security
switch OFF.
Reference Weight too small. The
weight on the platform is too small to
define a valid reference weight.
GlobalTestSupply
www.
Defender 6000 Indicators
Break the seal or replace the original
base/indicator.
TM
Switch back the original EasyConnect
base.
1.
Check if the cable connection is ok.
2.
Break the seal, perform factory reset
and recalibrate the scale.
TM
1.
Break the seal or replace the original
base/indicator.
2.
Service required.
Break the seal, connect base to the indicator
and power on the scale. Power off the scale
and perform sealing again.
Calibrate scale.
Replace batteries (i-DT61PW).
Use correct calibration weight.
Refer to LEGAL FOR TRADE chapter for
details about how to set the security switch to
the ON position.
Use a heavier weight for sample.