INSTRUMENT PROMPT MESSAGE
The instrument is extremely stable and reliable so, generally speaking, errors are rather uncommon. Once
an error occurs, first clear the error then observe whether the instrument still has the error after the power-
on. Do not rush to repair the scale body or instrument. Repair the instrument on the basis of the error code
if possible.
No.
Symbol
1
[ _EEE ]
[ EEE ]
┌----┐
2
[
]
3
└-----┘
[
]
┌-
-┐
4
[
no
]
└-
-┘
[
no
]
5
[--no--]
6
[Err 03]
7
[Err 05]
8
[Err 06]
9
[Err 07]
10
[Err 08]
11
[Err 09]
12
[Load]
13
[SETUP
14
[End ]
15
[Add--]
16
[- OVER -]
17
[Ld---]
Description
Unable to reset after start-up
The weighed object has been
over full range for 9 days
The weighed object has been
below 0 for 5 days
Out of the resetting range
Invalid operation
EEPROM checksum and error Press [ON/OFF] to reprint the factory
The calibration input weight is
too low
The weight being calibrated is
too light
The scale is in dynamic mode
Date and time setting error
AD initialisation error
When loading the scale, it
indicates the loading weight
Enter into the menu setting
End of zero point and loading
point calibration
Including current display weight
in accumulated value
Accumulated weight beyond
threshold
Loading default value
Action
1. Make sure the scale is in no-load
state at start-up;
2. Perform zero calibration again.
Reduce the weight on the weighing
platform
Press [Zero] to reset
Check whether the weighing platform
has any weight.
value. Start the scale up again. If the
message still appears, return the
product to the factory for repairs.
Calibrate the scale again if the
situation does not occur;
Attention: this manual includes all
parameters of the factory's
instruments.
Input is ≥10 % of the full range weight
Load is ≥10% of the full range weight
Inspect the scale body
Set the proper date and time
If the error occurs after the restart,
return the scale to the factory for
repairs
Load weight according to
requirements
Press [ON/OFF] to continue setting.
Clear accumulated weight in time
16