Herunterladen Diese Seite drucken

acumed Nail 2 System Gebrauchsanweisung Seite 48

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 21
EN
Instructions for use
POST-STERILIZATION INSPECTION
• Do not store or use sterile devices if they are not dry.
- Moisture supports the survival of microorganisms.
- Moisture remaining on wrapped or contained products after sterilization could compromise the
sterile barrier.
- Moisture can corrode metal and dull sharp edges.
• Inspect the sterile barrier for signs of damage. Do not use the product if the sterile barrier has been
compromised.
STORAGE CONDITIONS
STORAGE OF SETS AFTER STEAM STERILIZATION
• Items should be stored under controlled conditions in a manner that minimizes the potential for
contamination per ANSI/AAMI ST79:2017. Refer to sterilization wrap or rigid container manufacturer's
IFU for limits on sterile product storage time and storage requirements for temperature and humidity.
STORAGE OF PACKAGED NON-STERILE AND STERILE PRODUCT
• Final packaged product should be stored at room temperature (59-77
direct sunlight, pests, and high humidity.
English (non-U.S.) – EN / PAGE 48
o
F or 15-25
o
C) and protected from