3. When the «°C» or «°F» icon is flashing, a beep sound is
heard and the thermometer is ready for measurement 8.
4. Place the probe gently onto the area approximately 1cm
above the center of the eyebrow. If the eyebrow area is
covered with hair, sweat or dirt, please remove the
obstacle beforehand to improve the reading accuracy.
5. Press the START button once and release it. Start to
move the instrument steadily from 1 cm above the center
of the eyebrow to the temple area. A long beep sound
will emit after 3 seconds to acknowledge the completion
of measurement. If the probe movement is too slow to
reach the temple area before the long beep sound is
heard, repeat the measurement as described above, but
simply move the probe a little quicker this time.
6. Read the recorded temperature from the LCD display.
NOTE:
• Patients and thermometer should stay in steady-
state room condition for at least 30 minutes.
• Don't take a measurement while or immediately after
nursing a baby.
• Don't use the forehead thermometer in high humidity
environments.
FR 100
• Patients should not drink, eat, or exercise before/while
taking the measurement.
• Before placing the thermometer probe onto the
measurement area, remove dirt, hair, or sweat.
• Don't move the measurement device from the measuring
area before hearing the termination beep.
• Use an alcohol swab to carefully clean the probe and
wait for 15 minutes before taking a measurement on
another patient.
• 10 short beeps and a red LCD backlight alert the patient
that he/she may have a temperature equal to or higher
than 37.5 °C.
• Always take the temperature in the same location, since
temperature readings may vary according to location.
• In the following situations it is recommended that three
temperatures in the same area be taken and the highest
one taken as the reading:
1. New born infants in the first 100 days.
2. Children under three years of age with a compromised
immune system and for whom the presence or
absence of fever is critical.
3. When the user is learning how to use the thermometer
for the first time until he/she has familiarized himself/
herself with the instrument and obtains consistent
readings.
4. If the measurement is surprisingly low.

6. Changing between Celsius and Fahrenheit

This thermometer can display temperature measurements in
either Fahrenheit or Celsius. To switch the display between
°C and °F, simply turn OFF the unit, press and hold the
START button 2 for 5 seconds; after 5 seconds, the current
measurement scale («°C» or «°F» icon) will be flashing on the
display AK. Change the measurement scale between °C and
°F by pressing the START button 2. When the measurement
scale has been chosen, wait for 5 seconds and the unit will
automatically enter the «ready for measuring» mode.